Certain reactions need to be run in an inert environment with no O2 or water vapor present.
In most cases, flushing the reaction vessel with an inert gas, such as N2 or argon is sufficient to eliminate
unwanted interaction with the atmosphere. This page describes how to flush a vessel to run a reaction under
an inert gas environment.
- balance
- weighing boat | paper
- glass vial | plastic > 5mL
- glass | plastic transfer pipette
- Teflon film (~20um thick)
- spin casting head
- heated centrifuge | spinning apparatus
- dry-ice | liquid nitrogen
- Tefzel tape (DuPont)
- 15 cm plastic | wood | metal rod for hanging LCE
This material is prepared using a heated centrifuge in a spin casting head. Before
starting material preparation, pre-heat the spin casting head to 60 ~ 65oC.
The first step is preparation of the catalyst solution and reaction solution
- The platinum catalyst solution is prepared as a 1 to 2% weight/weight (w/w) solution
in methylene chloride. Typically, this solution is prepared in quantities of 100 to 200uL in a brown vial and stored
in the refrigerator.
- Example calculation for 1% (w/w)solution: Add 2 mg of the platinum
catalyst to brown glass vial. Add methylene
chloride drop wise until the total mass is 100 mg. This will be enough to prepare 5 batches of LCE
Glass vial used in preparation of catalyst solution. The white crystalline powder
at the front of the vial is the Pt catalyst crystals
- Dry the flask in an oven or with a heat gun and allow to return to
room temperature slowly so that condensation does not form on the
surface of the glass. Add all required solvents, chemicals, or
stir bars to the flask, and place rubber septa over all joints
(openings) in the flask.
- Secure the flask with a clamp.
- The picture to the right shows a needle-syringe-hose assembly that
is inserted through the rubber septum on the reaction flask. Cut
off the 'handle' section at the end of the syringe and insert into the
supply and exhaust hoses.
- Connect the exhaust hose to the bubbler ( if available), and the
other hose to nitrogen supply source, either a manifold or gas supply tank.
- Gently push the syringe's needle for the exhaust and supply lines
through the rubber septum. It is not a problem if there is only one
septum on the flask. When you are finished, the entire assembly
should be similar to the system shown in the picture on the right.
- Flush the flask with gas for 2 to 3 minutes. If you have a
bubbler, a flow rate of a bubble every 2 to 3 seconds is sufficient.
Flushing the flask for too long a period of time or at too high a flow
rate will drive on solvent and is not recommened.
- Place the plastic covers back over the needle tips. Be careful not
to launch the plastic covers the next time you run a reaction
under inert gas. For a picture of a bubbler, please see the image at the
end of the word or pdf document